15
Technical specifications
15.1
MADSEN Xeta
Type identification
MADSEN Xeta is type 1067 from Natus Medical Denmark ApS.
Channels
2 separate and identical channels
Pure tone frequencies
AC:
11 standard 125 - 8000 Hz
BC:
250 - 8000 Hz standard frequencies
Insert phones
125 - 8000 Hz standard frequencies
Accuracy:
Better than 1 %.
Modulation
FM (Warble):
1 - 20 Hz in 1 Hz steps. Mod. width 1% - 25% in 1% steps
AM for SISI:
5, 4, 3, 1, 0.75, 0.50, 0.25 dB HL steps
Attenuator
5 dB HL step resolution over the entire range
Attenuator accuracy
In whole range:
better than 3 dB HL
Between two consecutive attenuator positions:
5 dB HL step:
better than 1 dB HL
HL Range
Maximum output will be limited by the transducer.
AC:
-10 to 120 dB HL at mid-frequencies
BC:
-10 to 70 dB HL at mid-frequencies
Masking
Narrow band noise
Total harmonic distortion
Air < 2.5 %
Bone < 5 %
Otometrics - MADSEN Xeta
95
Summary of Contents for Madsen Xeta
Page 1: ...MADSEN Xeta Reference Manual Doc No 7 50 0870 EN 17 Part No 7 50 08700 EN ...
Page 6: ...Index 103 6 Otometrics MADSEN Xeta ...
Page 26: ...3 Overview of MADSEN Xeta 26 Otometrics MADSEN Xeta ...
Page 44: ...4 Navigating in the OTOsuite Audiometry Module 44 Otometrics MADSEN Xeta ...
Page 54: ...5 Preparing for testing 54 Otometrics MADSEN Xeta ...
Page 66: ...6 MADSEN Xeta Test Settings 66 Otometrics MADSEN Xeta ...
Page 86: ...11 Unpacking and installing 86 Otometrics MADSEN Xeta ...
Page 88: ...12 Maximum non destructive voltage 88 Otometrics MADSEN Xeta ...
Page 94: ...14 Standards and safety 94 Otometrics MADSEN Xeta ...
Page 102: ...15 Technical specifications 102 Otometrics MADSEN Xeta ...