Default starting fre-
quency
1000 Hz.
You cannot change this value.
If this frequency is not successfully tested, the entire test is aborted. Reinstruct the
patient and try again.
AS Frequencies
•
Use the
Frequency
knob to scroll between frequencies.
•
Use the
Right level
knob to set to
On
or
Off
. 1000 Hz is always
On
and cannot
be changed.
AS Familiarize
Use the
Right level
knob to select.
At the start of each frequency a clearly audible tone is presented as an example of
the presented signal. During the actual test, the signal to be presented is reduced
to the predetermined screening level.
Default: 40 dB HL
Range: 40 dB to 60 dB HL or Off
AS Method
This setting defines the number of threshold passages that determines the
threshold.
Use the
Right level
knob to set the threshold detection criterion to:
1 of 1, or
2 of 3
AS Level
Use the
Right level
knob to change the starting hearing level.
Min. 0 dB, max. 40 dB HL.
AS Do AT when AS fails
Use the
Right level
knob to select/deselect.
•
On
To include Auto Threshold in the Auto Screening test.
The Auto Threshold test is activated immediately when a frequency fails. The
test returns to Auto Screening when a threshold has been determined for the
specific frequency.
•
Off
AS AT Criterion
This setting defines the number of threshold passages that determines the
threshold. This applies when Auto Threshold is used as a complimentary function at
frequencies where the screening fails. See also the setting
, to include Auto Threshold in your Auto Screening test.
Use the
Right level
knob to set the threshold detection criterion to:
1 of 1
2 of 3, or
3 of 5
Otometrics - MADSEN Xeta
57
6 MADSEN Xeta Test Settings
Summary of Contents for Madsen Xeta
Page 1: ...MADSEN Xeta Reference Manual Doc No 7 50 0870 EN 17 Part No 7 50 08700 EN ...
Page 6: ...Index 103 6 Otometrics MADSEN Xeta ...
Page 26: ...3 Overview of MADSEN Xeta 26 Otometrics MADSEN Xeta ...
Page 44: ...4 Navigating in the OTOsuite Audiometry Module 44 Otometrics MADSEN Xeta ...
Page 54: ...5 Preparing for testing 54 Otometrics MADSEN Xeta ...
Page 66: ...6 MADSEN Xeta Test Settings 66 Otometrics MADSEN Xeta ...
Page 86: ...11 Unpacking and installing 86 Otometrics MADSEN Xeta ...
Page 88: ...12 Maximum non destructive voltage 88 Otometrics MADSEN Xeta ...
Page 94: ...14 Standards and safety 94 Otometrics MADSEN Xeta ...
Page 102: ...15 Technical specifications 102 Otometrics MADSEN Xeta ...