Start
to start the test.
If the patient fails to respond to the presented signal, the level is automatically increased until the patient
responds to the signal.
The signal is by default presented for each tone in the following order:
1000 Hz, 2000 Hz, 4000 Hz, and 8000 Hz, followed by 1000 Hz (if Cross Check is enabled), 500 Hz and 250 Hz.
If the patient responds before the signal is presented, the message
Illegal resp.!
flashes on the display and the
test is paused.
Note
•
You can use
Talk Over
at any time during the test. This will pause the test. The test is resumed when you
release the
Talk Over
button.
3. MADSEN Xeta beeps twice and the display shows that the test is completed. The results are stored automatically.
4. If the Masking Assistant flashes M, alternately ?, to indicate that masking may be recommended for specific fre-
quencies, scroll through the frequencies. The audiogram symbol will flash at the specific frequencies where masking
may be recommended.
5. To retest any frequencies, see
Air Conduction threshold test (manual)
and, if required,
6. If required, select the next patient record (press
Previous
/
Next
) or press
New Patient
.
2.3.2
Automatic Screening
Screening audiometry is a quick and simple way to identify patients in need of thorough audiological evaluation.
The test involves presenting tones at a predefined Hearing Level, the screening level. The patient is either able or unable
to hear the presented tone. The results show whether the Hearing Thresholds are better or worse than the defined
screening level.
Screening audiometry can be combined with Threshold determination of the frequencies where the screening level is not
audible.
1. Press
Auto Screening
. The button lights to indicate that you have selected the test.
A.
Auto Screening
B.
Current frequency
C.
Starting Hearing Level
D.
Starting ear selected
For symbols used, see
.
2. Press
Start
or
Setup
.
–
Start
to start the test.
–
Setup
to change the specific Auto Screening test settings (see
).
Otometrics - MADSEN Xeta
13
2 Getting started
Summary of Contents for Madsen Xeta
Page 1: ...MADSEN Xeta Reference Manual Doc No 7 50 0870 EN 17 Part No 7 50 08700 EN ...
Page 6: ...Index 103 6 Otometrics MADSEN Xeta ...
Page 26: ...3 Overview of MADSEN Xeta 26 Otometrics MADSEN Xeta ...
Page 44: ...4 Navigating in the OTOsuite Audiometry Module 44 Otometrics MADSEN Xeta ...
Page 54: ...5 Preparing for testing 54 Otometrics MADSEN Xeta ...
Page 66: ...6 MADSEN Xeta Test Settings 66 Otometrics MADSEN Xeta ...
Page 86: ...11 Unpacking and installing 86 Otometrics MADSEN Xeta ...
Page 88: ...12 Maximum non destructive voltage 88 Otometrics MADSEN Xeta ...
Page 94: ...14 Standards and safety 94 Otometrics MADSEN Xeta ...
Page 102: ...15 Technical specifications 102 Otometrics MADSEN Xeta ...