9. Continue to use the right and/or left
Level
knob to increase/decrease intensities until you
have determined all thresholds.
10. Continue to press
Store
after each threshold is found to store that particular threshold.
11. After each threshold is determined and stored, you will notice that as you turn the
Level
knob to scroll frequencies, each stored threshold at that particular frequency will be
marked by an asterisk (*) and the relevant audiogram symbol indicating the stored
threshold.
12. If you are not performing the test online with the OTOsuite Audiometry Module, plot the
results in an audiogram sheet.
13. Repeat this procedure for the other ear.
Note
•
If the difference between the air conducted threshold and the unmasked bone con-
duction threshold of the same ear is 15 dB or greater, it is recommended that you retest
bone conduction thresholds with masking applied to the contralateral ear. See
14. If required, select the next patient record (press
Previous
/
Next
) or press
New Patient
.
7.5
Testing with masking
See
•
Preparing for testing with masking
,
and
•
.
7.5.1
Air conduction with masking - suggested procedure
Prerequisites
The unmasked hearing thresholds have first been determined as described in
Air Conduction threshold test (manual)
.
Prerequisites as in
Air Conduction threshold test (manual)
.
•
Prepare the patient for testing with masking. See
Preparing for testing with masking
.
•
It is recommended to use the default setup used in connection with selecting
Contra Air
and
Mono Insert
for present-
ing the masking signal as continuous presentation.
This means that you will not be using the Interrupter button to control the masking signal.
72
Otometrics - MADSEN Xeta
7 Testing a patient with MADSEN Xeta
Summary of Contents for Madsen Xeta
Page 1: ...MADSEN Xeta Reference Manual Doc No 7 50 0870 EN 17 Part No 7 50 08700 EN ...
Page 6: ...Index 103 6 Otometrics MADSEN Xeta ...
Page 26: ...3 Overview of MADSEN Xeta 26 Otometrics MADSEN Xeta ...
Page 44: ...4 Navigating in the OTOsuite Audiometry Module 44 Otometrics MADSEN Xeta ...
Page 54: ...5 Preparing for testing 54 Otometrics MADSEN Xeta ...
Page 66: ...6 MADSEN Xeta Test Settings 66 Otometrics MADSEN Xeta ...
Page 86: ...11 Unpacking and installing 86 Otometrics MADSEN Xeta ...
Page 88: ...12 Maximum non destructive voltage 88 Otometrics MADSEN Xeta ...
Page 94: ...14 Standards and safety 94 Otometrics MADSEN Xeta ...
Page 102: ...15 Technical specifications 102 Otometrics MADSEN Xeta ...