10
Maintenance and calibration
MADSEN Xeta requires no preventive maintenance except for regular calibration of the transducers.
See
10.1
Service and repair
It is recommended that you keep the packing material in which MADSEN Xeta was delivered. If you need to send it in for
service, the original packing material will ensure protection against damage during transport, etc.
Warning
•
For the sake of safety and in order not to void the warranty, service and repair of electro-medical equip-
ment should be carried out only by the equipment manufacturer or by service personnel at authorized workshops. In
case of any defects, make a detailed description of the defect(s) and contact your supplier. Do not use a defective
device.
Note
•
There are no user-serviceable parts inside the MADSEN Xeta cabinet.
10.1.1
Fuses
Warning
•
Risk of fire
. Before replacing a fuse, first switch off the instrument and disconnect it from the
mains power supply.
The fuse-holder is located above the AC power connector.
To remove a fuse:
•
Unscrew the end of the fuse-holder using a screwdriver with a blade at least 6 mm wide.
Fuse types
The fuse is a push-fit in the end of the fuse-holder. When you replace fuses, use only fuses of type T 1 A H/250 V, 5 mm x
20 mm.
10.2
Cleaning
Cleaning the device
There are no specific requirements to sterilization or disinfection of the device.
Make sure that the instrument is kept clean and free of dust:
•
Remove dust using a soft brush.
•
To clean the cabinet, use a soft, slightly damp cloth with a small amount of mild detergent on it.
Caution
•
Do not allow any moisture inside the instrument!
•
To clean the display, use a dry cloth or soft brush. Note that the display is laminated for maximum readability and must
therefore be treated with care.
Otometrics - MADSEN Xeta
81
Summary of Contents for Madsen Xeta
Page 1: ...MADSEN Xeta Reference Manual Doc No 7 50 0870 EN 17 Part No 7 50 08700 EN ...
Page 6: ...Index 103 6 Otometrics MADSEN Xeta ...
Page 26: ...3 Overview of MADSEN Xeta 26 Otometrics MADSEN Xeta ...
Page 44: ...4 Navigating in the OTOsuite Audiometry Module 44 Otometrics MADSEN Xeta ...
Page 54: ...5 Preparing for testing 54 Otometrics MADSEN Xeta ...
Page 66: ...6 MADSEN Xeta Test Settings 66 Otometrics MADSEN Xeta ...
Page 86: ...11 Unpacking and installing 86 Otometrics MADSEN Xeta ...
Page 88: ...12 Maximum non destructive voltage 88 Otometrics MADSEN Xeta ...
Page 94: ...14 Standards and safety 94 Otometrics MADSEN Xeta ...
Page 102: ...15 Technical specifications 102 Otometrics MADSEN Xeta ...