REVERSE
The masking signal follows the masking interrupter. The masking signal disappears
when button is pressed.
TOGGLE
When using toggled masking presentation, the masking signal changes from being
activated or deactivated each time the masking interrupter is pressed and
released. Also, when using toggled masking presentation, the masking signal is
deactivated as soon as the stimulus type (AC/BC), frequency, ear or patient is
changed. This allows for setting the appropriate masking level before the masking
signal is presented to the patient.
Masking Assistant
Mask. assistant
Enables indication of recommended masking (see description in
).
Default:
On
Extended range
Extended Range
Selection of one of two modes (see description in
):
•
Auto
•
Manual
Default:
Auto
Display
Display
Controls display setup. Fixes one of the four left/right configurations:
SIGNAL AS L/R
The signal for the left ear is shown on the left side, and that for the right ear on
the right side. Masking is on the opposite side.
SIGNAL AS R/L
The signal for the left ear is shown on the right side, and that for the right ear on
the left side. Masking is on the opposite side.
FIX SIGNAL ON L
The signal is shown on the left side, and masking on the right side.
FIX SIGNAL ON R
The signal is shown on the right side, and masking on the left side.
Otometrics - MADSEN Xeta
61
6 MADSEN Xeta Test Settings
Summary of Contents for Madsen Xeta
Page 1: ...MADSEN Xeta Reference Manual Doc No 7 50 0870 EN 17 Part No 7 50 08700 EN ...
Page 6: ...Index 103 6 Otometrics MADSEN Xeta ...
Page 26: ...3 Overview of MADSEN Xeta 26 Otometrics MADSEN Xeta ...
Page 44: ...4 Navigating in the OTOsuite Audiometry Module 44 Otometrics MADSEN Xeta ...
Page 54: ...5 Preparing for testing 54 Otometrics MADSEN Xeta ...
Page 66: ...6 MADSEN Xeta Test Settings 66 Otometrics MADSEN Xeta ...
Page 86: ...11 Unpacking and installing 86 Otometrics MADSEN Xeta ...
Page 88: ...12 Maximum non destructive voltage 88 Otometrics MADSEN Xeta ...
Page 94: ...14 Standards and safety 94 Otometrics MADSEN Xeta ...
Page 102: ...15 Technical specifications 102 Otometrics MADSEN Xeta ...