Set the initial level of the presented signal to an audible level: Turn the
Level
knob to
40 dB HL.
This is to familiarize the patient with the test signal. Other familiarization procedures may be
preferred.
•
If you intend to test both ears, test the better ear first. If this is unknown, start with the right
ear. Press
L<-->R
to select the ear you wish to test.
Testing
1. Press the
Present
button (right or left) to present the tone to the patient for 1 to 1.5
seconds.
2. If the patient does not respond to the presented signal, turn the
Level
knob to increase
the signal level in steps of 10 dB until the patient responds.
3. Reduce the signal level by 20 dB until the patient does not respond to the signal.
4. Increase the signal level in steps of 5 dB and present the signal to the patient at each level
until the patient responds to the signal.
5. Reduce the signal level by 10 dB and present signals at levels increased by 5 dB until the
patient responds.
6. Repeat item
, 1 to 3 times until the patient has responded 3 times at the same level.
This level represents the unmasked hearing threshold at 1000 Hz based on 3-5 ascending
series.
In some situations, it may be satisfactory with two responses at the same level.
If the patient has responded less than 3 times at the same level after 5 series, present a
signal at 10 dB above the level where a response was last registered.
Then repeat items
to
7. When the threshold is determined, press
Store
to store that particular threshold in the
selected patient record.
8. Turn the
Frequency
knob to select the next frequency.
You can link the Store and Next Frequency functions, so that the test progresses with the
next frequency when you press
Store
. See
Otometrics - MADSEN Xeta
71
7 Testing a patient with MADSEN Xeta
Summary of Contents for Madsen Xeta
Page 1: ...MADSEN Xeta Reference Manual Doc No 7 50 0870 EN 17 Part No 7 50 08700 EN ...
Page 6: ...Index 103 6 Otometrics MADSEN Xeta ...
Page 26: ...3 Overview of MADSEN Xeta 26 Otometrics MADSEN Xeta ...
Page 44: ...4 Navigating in the OTOsuite Audiometry Module 44 Otometrics MADSEN Xeta ...
Page 54: ...5 Preparing for testing 54 Otometrics MADSEN Xeta ...
Page 66: ...6 MADSEN Xeta Test Settings 66 Otometrics MADSEN Xeta ...
Page 86: ...11 Unpacking and installing 86 Otometrics MADSEN Xeta ...
Page 88: ...12 Maximum non destructive voltage 88 Otometrics MADSEN Xeta ...
Page 94: ...14 Standards and safety 94 Otometrics MADSEN Xeta ...
Page 102: ...15 Technical specifications 102 Otometrics MADSEN Xeta ...