12
Maximum non-destructive voltage
12.1
Input and output sockets
Maximum non-destructive input voltages for input and output sockets relative to earth connection when the instrument is
switched on.
Mono insert socket
10 V AC
±10 V DC
AC left and right socket
10 V AC
±10 V DC
BC socket
10 V AC
±10 V DC
Patient Mic/Talk Back socket
12 V AC
±12 V DC
Patient responder socket
12 V AC
±12 V DC
AC mains inlet
1500 V AC
±370 V DC
12.2
RS232 interface sockets
pin 1 (nc)
240 V AC
±240 V DC
pin 2 (RxDout)
15 V AC
*±15 V DC
pin 3 (TxDin)
30 V AC
*±30 V DC
pin 4 (DTR)
30 V AC
*±30 V DC
pin 5 (ISO Gnd.)
30 V AC
±30 V DC
pin 6 (DSR)
30 V AC
*±30 V DC
pin 7 (RTSin)
30 V AC
*±30 V DC
pin 8 (CTSout)
15 V AC
*±15 V DC
pin 9 (nc)
240 V AC
±240 V DC
Relative to pin 5 on the RS232 interface socket.
Otometrics - MADSEN Xeta
87
Summary of Contents for Madsen Xeta
Page 1: ...MADSEN Xeta Reference Manual Doc No 7 50 0870 EN 17 Part No 7 50 08700 EN ...
Page 6: ...Index 103 6 Otometrics MADSEN Xeta ...
Page 26: ...3 Overview of MADSEN Xeta 26 Otometrics MADSEN Xeta ...
Page 44: ...4 Navigating in the OTOsuite Audiometry Module 44 Otometrics MADSEN Xeta ...
Page 54: ...5 Preparing for testing 54 Otometrics MADSEN Xeta ...
Page 66: ...6 MADSEN Xeta Test Settings 66 Otometrics MADSEN Xeta ...
Page 86: ...11 Unpacking and installing 86 Otometrics MADSEN Xeta ...
Page 88: ...12 Maximum non destructive voltage 88 Otometrics MADSEN Xeta ...
Page 94: ...14 Standards and safety 94 Otometrics MADSEN Xeta ...
Page 102: ...15 Technical specifications 102 Otometrics MADSEN Xeta ...