Storing results
1. When the test is completed for a frequency, press
Store
to store:
–
SISI % score, number of dB increments administered,
–
dB increment,
–
dB level for test signal and masking signal (if applicable).
–
An * symbol appears by the frequency and an x/o symbol for the left
or right side.
2. Test the next frequency.
Pausing the test
•
If you press
Store
while the test is in progress, the test will be stopped and saved with the current number of dB incre-
ments.
Deleting specific values
•
If you hold
Store
at a frequency at which data are stored, the values for that frequency will be deleted.
7.7
ABLB (Alternate Binaural Loudness Balance)
Note
•
This test is optional.
In this test a tone is transmitted at the same frequency to the left and right ear
alternately
. The patient must decide when
the same volume is achieved in both ears. The patient indicates this by pressing the patient responder.
1. To select the test press
More
once or more until ABLB is shown.
2. Press
ABLB
once, and the test display is shown.
The top line shows the intensity for Right and Left channel and the test
name.
The frequency of the test tone and the rate of alternation from ear to
ear and back again (the alternating frequency) appear in the bottom line.
Changing alternating frequency
To change the alternating frequency:
1. Press
Setup
once and use the
Right level
knob (0.25 - 2.5 Hz).
2. Press
Setup
again to implement the new setting.
3. You can present the signal in two ways:
–
present the signal to the patient by pressing the
Present Signal
button, or
–
press the
Reverse
button to present the signal continuously. You can then interrupt the test signal by pressing
the
Present Signal
button.
Signal presentation (normal/
Reverse
) follows the normal rules (see
).
76
Otometrics - MADSEN Xeta
7 Testing a patient with MADSEN Xeta
Summary of Contents for Madsen Xeta
Page 1: ...MADSEN Xeta Reference Manual Doc No 7 50 0870 EN 17 Part No 7 50 08700 EN ...
Page 6: ...Index 103 6 Otometrics MADSEN Xeta ...
Page 26: ...3 Overview of MADSEN Xeta 26 Otometrics MADSEN Xeta ...
Page 44: ...4 Navigating in the OTOsuite Audiometry Module 44 Otometrics MADSEN Xeta ...
Page 54: ...5 Preparing for testing 54 Otometrics MADSEN Xeta ...
Page 66: ...6 MADSEN Xeta Test Settings 66 Otometrics MADSEN Xeta ...
Page 86: ...11 Unpacking and installing 86 Otometrics MADSEN Xeta ...
Page 88: ...12 Maximum non destructive voltage 88 Otometrics MADSEN Xeta ...
Page 94: ...14 Standards and safety 94 Otometrics MADSEN Xeta ...
Page 102: ...15 Technical specifications 102 Otometrics MADSEN Xeta ...