7.7.1
Using STORE in the ABLB (Fowler) test
Storing equal volume values
Press
Store
to store values of "equal volume".
•
You can store up to five sets of measurements for each test frequency.
If you try to save more than five sets of measurements for a given frequency, the message "No more STORE" is shown.
•
You can store a maximum of four test frequencies, which you can select.
If you have stored two or more measurements at the same frequency, you can turn the frequency knob to step
through each of these measurements in turn before changing to the next frequency.
Deleting individual measurements
•
To delete an individual measurement, press and hold
Store
while the measurement is displayed. Other measurements
at the same frequency and at other frequencies remain unaffected.
If you start a test, and data is stored, you cannot just change the signal type. If you do, the system will not save the data.
The message "Not allowed" is displayed and the button(s) that has/have been wrongly set flash in the correct position
(three short flashes).
Masking
Note
•
The audiometer cannot perform masking during this test.
7.8
The Stenger test
Note
•
This test is optional.
The Stenger test is used to reveal a unilateral non-organic hearing loss in one ear.
1. Before you perform a Stenger test, measure the patient's tone thresholds for the left and right ears.
2. To select the Stenger test press
More
(one or more times) until "STEN" appears.
3. Select the test by pressing
STEN
once. The following display appears:
A tone is transmitted at the same frequency on the left and right sides simultaneously.
4. Press either of the
Present Signal
keys to present the tone to the patient.
Changing dB level
5. You can change the dB level setting separately for the right and the left ear: Set the dB level to 10 dB above the
threshold in the better ear, and 10 dB below the threshold in the poorer ear.
6. You can present the signal in two ways:
–
present the signal to the patient by pressing the
Present Signal
button, or
Otometrics - MADSEN Xeta
77
7 Testing a patient with MADSEN Xeta
Summary of Contents for Madsen Xeta
Page 1: ...MADSEN Xeta Reference Manual Doc No 7 50 0870 EN 17 Part No 7 50 08700 EN ...
Page 6: ...Index 103 6 Otometrics MADSEN Xeta ...
Page 26: ...3 Overview of MADSEN Xeta 26 Otometrics MADSEN Xeta ...
Page 44: ...4 Navigating in the OTOsuite Audiometry Module 44 Otometrics MADSEN Xeta ...
Page 54: ...5 Preparing for testing 54 Otometrics MADSEN Xeta ...
Page 66: ...6 MADSEN Xeta Test Settings 66 Otometrics MADSEN Xeta ...
Page 86: ...11 Unpacking and installing 86 Otometrics MADSEN Xeta ...
Page 88: ...12 Maximum non destructive voltage 88 Otometrics MADSEN Xeta ...
Page 94: ...14 Standards and safety 94 Otometrics MADSEN Xeta ...
Page 102: ...15 Technical specifications 102 Otometrics MADSEN Xeta ...