digiBASE-E ™ High-Performance Power-over-Ethernet Multichannel Analyzer / PMT Base for Scintillation Detectors
48
preset. The preset should always be set to a multiple of one second. See also
CLEAR_PRESETS and SHOW_TRUE_PRESET.
Execution Errors:
%131135083<CR>
The command was attempted while spectrum acquisition was
in progress. No action was taken.
SET_ULD chans
Sets the upper level discriminator to chans. which must be between 0 and [conversion
gain
!
1]. See also SHOW_ULD.
SET_VIEW value
Supports the digiBASE-E Routing gate mode. Switches the data view in the MAESTRO
spectrum window between the regular histogram memory (SET_VIEW 0) and the alternative
histogram memory (SET_VIEW 1). Note that this command affects only the spectrum view
for the local PC. Users attached to this digiBASE-E via other network nodes will, by default,
see the regular histogram memory data in MAESTRO.
SET_WINDOW [start, length]
Sets the window of interest to the specified start channel and number of channels. The
channels of spectral data in the window of interest are affected by commands such as
CLEAR, SET_DATA, and WRITE. If neither start nor length is provided, the window is set
to the maximum size allowed by the conversion gain specified. The window of interest is
always set to the maximum size after a SET_DEVICE or SET_SEGMENT command.
Execution Errors:
%131128085<CR>
The start channel was higher than the conversion gain.
%131129086<CR>
The length specified one or more channels that were too high
conversion gain.
%131132080<CR>
The start channel was specified without a length. If one
value is given, the other must also be given.
SET_ZERO_ADJUSTMENT value
Sets the zero stabilization adjustment to an arbitrary value from
!
65535 to 65535. The total
range of the adjustment value represents ±256 channels. This adjustment is usually only
made by the gain stabilizer, and reset to 0 with the INITIALIZE_ZERO_STABILIZATION
command. See also SHOW_ZERO_ADJUSTMENT.
SET_ZERO_CHANNEL chan
Sets the center channel for the stabilizer zero peak. If a zero channel is chosen such that the
beginning channel or ending channel would be below channel 0 or above the maximum
channel as determined by the conversion gain, the zero peak width is reduced until the peak
fits the device boundaries.
Summary of Contents for digiBASE-E
Page 6: ...vi...