5. FIRMWARE COMMANDS AND RESPONSES
47
SET_OUTPUT_LOW [output-num]
Sets the Change Sample output to the low level. The digiBASE-E has only one sample-
changer BNC, therefore, if you include a value for output-num, it must be zero (output-num
is provided for compatibility with other ORTEC instruments). See also STEP_OUTPUT.
SET_RADIX_BINARY
Not used; included for backward compatibility.
SET_ROI start_chan,number_of_chans
Sets the ROI flags for the specified channels. This command can be used multiple times to
set ROI flags without affecting previously set flags.
SET_SEGMENT
Not used; included for backward compatibility. This command executes SET_WINDOW,
resetting the window of interest to the maximum number of channels.
SET_SHAP_FLAT value
Sets the width of the flattop to value. Value is in microseconds, ranging from 0.04 to 2.00 in
steps of 0.04. See also SHOW_SHAP_FLAT.
SET_SHAP_RISE value
Sets the rise time to value. Value is in microseconds, ranging from 0.6 to 2.00 in steps
of 0.04. See also SHOW_SHAP_RISE.
SET_TRIG_SAMPLE setting
Selects the triggering source in InSight Mode. See LIST_TRIG_SAMP for legal trigger
sources.
SET_TRUE ticks
Sets the true-time counter to the specified number of ticks. The number represents true time
in units of 20 ms (50 ticks/sec). Normally this value is set by the digiBASE-E during data
acquisition. See also CLEAR_COUNTERS and SHOW_TRUE.
Execution Errors:
%131135083<CR>
The command was attempted while spectrum acquisition was
in progress. No action was taken.
SET_TRUE_PRESET ticks
Sets the true-time preset to the specified number of ticks (20 ms/tick). During data
acquisition when the true-time counter reaches the preset number of ticks, the preset is
complete and the acquisition is stopped. Setting a true-time preset to 0 ticks disables the