digiBASE-E ™ High-Performance Power-over-Ethernet Multichannel Analyzer / PMT Base for Scintillation Detectors
46
Execution Errors:
%131135083<CR>
The command was attempted while spectrum acquisition
wasin progress. No action was taken.
SET_LIVE_PRESET ticks
Sets the live-time preset to the specified number of ticks (20 ms/tick). During data
acquisition when the live-time counter reaches the preset number of ticks, the preset is
complete and the acquisition is stopped. Setting a live-time preset to 0 ticks disables the
preset. See also CLEAR_PRESETS and SHOW_LIVE_PRESET.
Execution Errors:
%131135083<CR>
The command was attempted while spectrum acquisition was
in progress. No action was taken.
SET_LLD chans
Sets the lower level discriminator to chan, which must be between 0 and [conversion
gain
!
1]. See also SHOW_LLD.
SET_MODE_LIST
Puts the unit in List mode as opposed to InSight or PHA Mode. In List mode, individual
peak values are stored along with the time the pulse arrived are stored and transferred to
memory. See SET_MODE_PHA and SET_MODE_SAMPLE.
SET_MODE_PHA
This command sets the MCB to pulse height analysis mode for collection of histogram data.
SET_MODE_SAMPLE
Starts the InSight Virtual Oscilloscope mode.
SET_OUTPUT value
This sends the value to the Change Sample output (B-OUT on the digiBASE-E).
SET_OUTPUT 1 sets the output voltage high; SET_OUTPUT 0 sets the output voltage low.
These commands are respectively equivalent to SET_OUTPUT_HIGH and
SET_OUTPUT_LOW. See also SHOW_OUTPUT. See also STEP_OUTPUT.
SET_OUTPUT_HIGH [output-num]
Sets the Change Sample output to the high level. The digiBASE-E has only one sample-
changer BNC, therefore, if you include a value for output-num, it must be zero (output-num
is provided for compatibility with other ORTEC instruments). See also STEP_OUTPUT.