5. FIRMWARE COMMANDS AND RESPONSES
41
ENABLE_REMOTE
No function in the digiBASE-E . This command has been included for backward
compatibility.
ENABLE_THRESHOLD_AUTOMATIC
Enables automatic determination of the positive and negative thresholds. See also
DISABLE_THRESHOLD_AUTO, SHOW_THRESHOLD_AUTO,
SET_THRESHOLD_NEGATIVE, and SET_THRESHOLD_POSITIVE.
ENABLE_THRESHOLD_SAMPLE
ENABLE_TRIGGER_SPECIAL
No function; included for backward compatibility.
ENABLE_ZERO_STABILIZATION
Enables the stabilization of the zero peak. See also DISABLE_ZERO_STABILIZATION,
SHOW_ZERO_STABILIZATION, and INITIALIZE_ZERO_STABILIZATION.
INIT
Resets the digiBASE-E to factory defaults.
INIT_GAIN_STABILIZATION
Resets the gain peak stabilization adjustment to 1.0 (no adjustment). This value is reported
as 0 by the SHOW_GAIN_ADJUSTMENT command. See also
SET_GAIN_ADJUSTMENT, ENABLE_GAIN STABILIZATION, and
DISABLE_GAIN_STABILIZATION.
INIT_ZERO_STABILIZATION
Resets the zero peak stabilization adjustment to 0 (no adjustment). This value is reported
as 0 by the SHOW_ZERO_ADJUSTMENT command. See also
SET_ZERO_ADJUSTMENT, ENABLE_ZERO_STABILIZATION, and
DISABLE_ZERO_STABILIZATION.
LIST_GAIN_ADJUST
Displays the range of gain-stabilizer adjustment that can be returned with the
SHOW_GAIN_ADJUST command. A value of -100 corresponds to the internal value of -
100 and +100 corresponds to 100.
Response:
GAIN_ADJU -100 100 –100 100
Summary of Contents for digiBASE-E