5. FIRMWARE COMMANDS AND RESPONSES
45
SET_GAIN_WIDTH chans
Sets the width in channels for the stabilizer gain peak. The gain width must be chosen such
that the beginning channel is no lower than channel 0 and the ending channel is no higher
than the maximum channel as determined by the conversion gain. The gain channel and
width must be set before gain stabilization can be enabled. The absolute minimum width for
the gain peak is 3 channels. See also SHOW_GAIN_WIDTH, SET_GAIN_CHANNEL and
SHOW_GAIN_CHANNEL.
Execution Errors:
%131128085<CR>
The specified number of channels would create a peak that
was less than the minimum (3-channel) width or would be
outside the device’s range.
%131136084<CR>
The command was attempted while gain stabilization was
enabled.
SET_GATE value
Sets the gate to a value from 0 to 7, according to the gate order displayed with LIST_GATE.
See also SHOW_GATE. The gate modes are discussed in Section 2.2. See also
SHOW_GATE and LIST_GATE.
Legal Commands:
SET_GATE 0<CR>
Gate modeNone.
SET_GATE 1<CR>
Gate mode AcqCntrol.
SET_GATE 2<CR>
Gate mode AcqTrigger.
SET_GATE 3<CR>
Gate mode CoincGate.
SET_GATE 4<CR>
Gate mode EventCounter.
SET_GATE 5<CR>
Gate mode Routing.
SET_GATE 6<CR>
Gate mode Sync.
SET_GATE 7<CR>
Gate mode SyncMaster.
SET_HV value
This sets the HV bias to value, in volts, and stores value as the target HV.
SET_LENGTH_SAMPLE length
Sets the number of points in the Insight mode display to the specified value. Normally this
setting would remain at 1000.
SET_LIVE ticks
Sets the live-time counter to the specified number of ticks. The number represents live time
in units of 20 ms (50 ticks/s). Normally this value is set by the digiBASE-E during data
acquisition. See also CLEAR_COUNTERS and SHOW_LIVE.