digiBASE-E ™ High-Performance Power-over-Ethernet Multichannel Analyzer / PMT Base for Scintillation Detectors
44
SET_GAIN_ADJUSTMENT value
Sets the gain stabilization adjustment to an arbitrary value from
!
65535 to 65535. This
adjustment is usually made only by the gain stabilizer, and reset to 0 with the
INITIALIZE_GAIN_STABILIZATION command. See also
SHOW_GAIN_ADJUSTMENT.
Execution Errors:
%131134082<CR>
The command was attempted when the currently
selected device was other than device 1.
SET_GAIN_CHANNEL chan
Sets the center channel for the stabilizer gain peak. If a gain channel is chosen such that the
beginning channel or ending channel would be below channel 0 or above the maximum
channel as determined by the conversion gain, the gain peak width is reduced until the peak
fits the device boundaries.
Execution Errors:
%131128085<CR>
The specified channel number would create a peak
that was less than the minimum width (3 channels) or
would be outside the device’s range.
%131136084<CR>
The command was attempted while gain stabilization
was enabled.
SET_GAIN_CONV value
Sets the conversion gain. The conversion gain defines the number of channels within the
device that will used for spectral data. See also SHOW_GAIN_CONVERSION.
Legal Commands:
SET_GAIN_CONVERSION 0<CR>
Conversion gain set to default (2048).
SET_GAIN_CONVERSION 256<CR>
Conversion gain set to 256 channels.
SET_GAIN_CONVERSION 1024<CR>
Conversion gain set to 1024 channels.
SET_GAIN_FINE value
Sets the fine gain to value. Value is a floating point value from 0.3333 to 1.0000. Note that
this command does not reflect the setting of the coarse-gain jumper discussed in Appendix A.
See also SHOW_GAIN_FINE.
Legal Commands:
SET_GAIN_FINE 0.55 <CR>
Fine gain set to 0.55.
SET_GAIN_POSITIVE
Sets the amplifier input polarity to positive.
SET_GAIN_PRESET count
No function in the digiBASE-E; included for backward compatibility.
Summary of Contents for digiBASE-E
Page 6: ...vi...