5. FIRMWARE COMMANDS AND RESPONSES
49
Execution Errors:
%131128085<CR>
The specified channel number would create a peak that was
less than the minimum width (3 channels) or would be
outside the device’s range.
%131136084<CR>
The command was attempted while zero stabilization was
enabled.
SET_ZERO_WIDTH chans
Sets the width in channels for the stabilizer zero peak. The width must be chosen such that
the beginning channel is no lower than channel 0 and the ending channel is no higher than
the maximum channel as determined by the conversion gain. The zero channel and width
must be set before gain stabilization can be enabled. The absolute minimum width for the
peak is 3 channels. See also SHOW_ZERO_WIDTH, SET_ZERO_CHANNEL and
SHOW_ZERO_CHANNEL.
Execution Errors:
%131128085<CR>
The specified number of channels would create a peak that
was less than the minimum width (3 channels) or would be
outside the device’s range.
%131136084<CR>
The command was attempted while zero stabilization was
enabled.
SHOW_ACTIVE
Returns a 1 if the digiBASE-E is active (i.e., acquiring spectral data) or 0 if it is not active.
Responses:
$C00000087<CR>
Not active.
$C00001088<CR>
Active.
SHOW_BT
Returns the digiBASE-E’s board temperature in degrees Celsius.
Responses:
$F45.50<CR>
Board temperature is 45.5
E
C.
SHOW_CONFIGURATION
Returns a record that indicates the hardware configuration of the MCB. The record contains
information about the number of segments in an MCB device (always 1 for the
digiBASE-E), and the current conversion gain for each segment. The record is organized as
follows:
$J0204800001aaaaa00000[65 zeros here for total of 75 zeros]00000ccc
where aaaaa represents the conversion gain for the one and only segment in the currently
selected device, and ccc represents the record checksum. See Section 5.1.2 for more
information about response records and checksums.
Summary of Contents for digiBASE-E
Page 6: ...vi...