digiBASE-E ™ High-Performance Power-over-Ethernet Multichannel Analyzer / PMT Base for Scintillation Detectors
40
reinitialized with INITIALIZE_GAIN_STABILIZATION or SET_GAIN_PEAK. See also
SHOW_GAIN_STABILIZATION.
DISABLE_HV
Turns off the HV enable signal. See ENABLE_HV and SHOW_HV.
DISABLE_NAI
Switches the stabilizers to Ge mode. See ENABLE_NAI.
DISABLE_PUR
Turns off the pulse pileup rejector. See also ENABLE_PUR.
DISABLE_THRESHOLD_AUTOMATIC
DISABLE_THRESHOLD_SAMPLE
DISABLE_TRIGGER_SPECIAL
No function in the digiBASE-E. These commands have been included for backward
compatibility.
DISA_ZERO_STABILIZATION
Stops stabilization of the zero peak. The zero stabilization adjustment is held at its current
value until either zero stabilization is reenabled with the ENABLE_ZERO_
STABILIZATION command or reinitialized with INITIALIZE_ZERO_ STABILIZATION
or SET_ZERO_PEAK. See also SHOW_ZERO_STABILIZATION.
ENABLE_GAIN_STABILIZATION
Enables the stabilization of the gain peak. See also DISABLE_GAIN_STABILIZATION,
SHOW_GAIN_STABILIZATION, and INITIALIZE_GAIN_STABILIZATION.
ENAB_HV
Turns on the high-voltage enable output of the digiBASE-E and sets the HV to the target HV
value. See also DISABLE_HV and SHOW_HV.
Execution Errors:
%131137085<CR>
The high voltage could not be enabled due to a high voltage
shutdown condition.
ENABLE_NAI
Switches the stabilizers to NaI mode. The stabilizers use the “fine” adjustment in NaI mode
and the “very fine” adjustment in Ge mode. See DISABLE_NAI.
ENABLE_PUR
Turns on the pulse pileup rejector. See also DISABLE_PUR.
Summary of Contents for digiBASE-E
Page 6: ...vi...