4.3 Connection of an endoscope
39
CLV-S190 INSTRUCTION MANUAL
Ch.4
WARNING
• Do not look directly into the distal end of the endoscope, the distal end of the light
guide cable, or the output socket of the light source when they are emitting light.
The intense light may cause eye injury.
• Use compatible endoscopes only. Using an incompatible endoscope can result in
patient injury and/or equipment damage. Refer to the “
System chart” on page 93
for the compatible endoscope to be used with the light source.
• If the surface of the endoscope and light guide cable is dirty, reprocess it according
to the directions given in the endoscope’s instruction or reprocessing manual
before connecting it to the light source. Otherwise, it may cause patient injury,
equipment damage, and/or improper illumination.
• Since the light source irradiates strong examination light, the disconnected end of
the light guide cable or the distal end of the endoscope becomes very hot. To
prevent a fire hazard, do not bring the disconnected end of the light guide cable or
distal end of the endoscope in contact with a flammable object, such as operating
room drapes when the examination lamp is turned ON.
• Do not touch the distal end of the light guide connector of the endoscope, distal end
of the light guide cable, distal end of the light guide cable’s connector, or output
socket of the light source immediately after disconnecting it from the light source
because they are extremely hot. Operator or patient injury can result.
• Before connecting the endoscope connector to the light source, make sure that it is
completely dry. Otherwise, electric shock or equipment damage can result.
Connection of a rigid endoscope
WARNING
Connect equipment in the order described below. Otherwise, the light emitted from
the distal end of the light guide cable may cause operator and/or patient injury or a
fire by igniting flammable material, such as an operating room drape.
4.3
Connection of an endoscope
Summary of Contents for VISERA ELITE CLV-S190
Page 2: ......
Page 6: ...iv Contents CLV S190 INSTRUCTION MANUAL...
Page 20: ...14 Summary of the Functions of the Instrument CLV S190 INSTRUCTION MANUAL...
Page 22: ...16 1 1 Checking the package contents list CLV S190 INSTRUCTION MANUAL Ch 1...
Page 30: ...24 2 3 Rear and side panels CLV S190 INSTRUCTION MANUAL Ch 2...
Page 42: ...36 3 6 Connection to the AC mains power supply CLV S190 INSTRUCTION MANUAL Ch 3...
Page 62: ...56 4 11 After inspection CLV S190 INSTRUCTION MANUAL Ch 4...
Page 98: ...92 8 3 Returning the light source for repair CLV S190 INSTRUCTION MANUAL Ch 8...