34
3.6 Connection to the AC mains power supply
CLV-S190 INSTRUCTION MANUAL
Ch.3
DANGER
• Be sure to connect the power plug of the power cord directly to the mobile
workstation (WM-NP2, WM-NP1, or WM-WP1) or a grounded wall mains outlet. If
the light source is not grounded properly, it can cause an electric shock.
• Do not connect the power plug to a 2-pole power circuit with a 3-pole to 2-pole
adapter. It can prevent proper grounding and cause an electric shock.
• Do not connect the power plug using an extension cord. It can prevent proper
grounding and cause an electric shock.
WARNING
• Always keep the power plug dry. A wet power plug may cause electric shocks.
• Confirm that the hospital-grade wall mains outlet to which the light source is
connected has adequate electrical capacity that is larger than the total power
consumption of all connected equipment. If the capacity is insufficient, a fire can
result, or the circuit breaker may trip, and it may turn OFF not only the light source
but also all other equipment connected to the same power circuit.
• When using the mobile workstation, confirm that the mobile workstation has
adequate electrical capacity that is larger than the total power consumption of all
connected equipment. If the capacity is insufficient, drop in the supply voltage can
result or the electric protective device may trip and turn OFF all the equipment
connected to the mobile workstation.
• Do not bend, pull, or twist the power cord. Equipment damage including separation
of the power plug and disconnection of the cord wire as well as an electric shock
can result.
• Be sure to connect the power plug securely to prevent unintentional disconnection
during use. The equipment will not function.
• Do not extend a single wall mains outlet into multiple outlets for simultaneous
connection of both the electrosurgical unit and light source. Malfunction of the
equipment may result.
3.6
Connection to the AC mains power supply
Summary of Contents for VISERA ELITE CLV-S190
Page 2: ......
Page 6: ...iv Contents CLV S190 INSTRUCTION MANUAL...
Page 20: ...14 Summary of the Functions of the Instrument CLV S190 INSTRUCTION MANUAL...
Page 22: ...16 1 1 Checking the package contents list CLV S190 INSTRUCTION MANUAL Ch 1...
Page 30: ...24 2 3 Rear and side panels CLV S190 INSTRUCTION MANUAL Ch 2...
Page 42: ...36 3 6 Connection to the AC mains power supply CLV S190 INSTRUCTION MANUAL Ch 3...
Page 62: ...56 4 11 After inspection CLV S190 INSTRUCTION MANUAL Ch 4...
Page 98: ...92 8 3 Returning the light source for repair CLV S190 INSTRUCTION MANUAL Ch 8...