82
6.4 Lamp usage indicator reset
CLV-S190 INSTRUCTION MANUAL
Ch.6
1
Connect the power cord and press the power switch: the light source is turned ON.
2
Press the lamp button on the control panel for
1 second: the examination lamp is turned OFF,
and the “STBY” indicator lights up.
Figure 6.16
3
Press and hold the counter reset button on the
control panel for more than 3 seconds: the
lamp usage indicator is reset. Confirm that the
lamp usage indicator shows “0”.
Figure 6.17
NOTE
The counter reset button is inactive while the examination lamp is lighting. To reset
the lamp usage indicator, press the button while the examination lamp is in
standby.
4
Turn the light source OFF immediately.
5
Inspect the light source as described in Chapter 4, “Inspection” before use.
Lamp indicator “STBY”
Lamp button
Lamp usage indicator
Counter reset button
Counter reset button