3.3 Installation of equipment
27
CLV-S190 INSTRUCTION MANUAL
Ch.3
CAUTION
• Do not place any equipment other than the video system center on the top of the
light source. Equipment damage can result.
• Keep the ventilation grills of the light source clear. The ventilation grills are located
on the side and rear panels. Blockage can cause overheating and equipment
damage.
• Clean and vacuum dust the ventilation grills using a vacuum cleaner. Otherwise,
the light source may break down and gets damaged from overheating.
• Place the light source on a stable, level surface using the foot holders (MAJ-1205).
Otherwise, the light source may topple down or drop, and operator or patient injury
may occur, or equipment damage can result.
• When using the mobile workstation, confirm that excessive force is not applied to
the cord of the workstation. The cord may be disconnected, and malfunction can
result.
• If a trolley other than the mobile workstation (WM-NP2, WM-NP1, or WM-WP1) is
used, confirm that the trolley can withstand the weight of the equipment installed on
it.
• Do not install the light source in locations exposed to strong electromagnetic
radiation (e.g., in the vicinity of a microwave therapeutic equipment, MRI,
short-wave therapeutic equipment, radio equipment, or cellular/portable phones).
Otherwise, the light source may malfunction.
3.3
Installation of equipment
Summary of Contents for VISERA ELITE CLV-S190
Page 2: ......
Page 6: ...iv Contents CLV S190 INSTRUCTION MANUAL...
Page 20: ...14 Summary of the Functions of the Instrument CLV S190 INSTRUCTION MANUAL...
Page 22: ...16 1 1 Checking the package contents list CLV S190 INSTRUCTION MANUAL Ch 1...
Page 30: ...24 2 3 Rear and side panels CLV S190 INSTRUCTION MANUAL Ch 2...
Page 42: ...36 3 6 Connection to the AC mains power supply CLV S190 INSTRUCTION MANUAL Ch 3...
Page 62: ...56 4 11 After inspection CLV S190 INSTRUCTION MANUAL Ch 4...
Page 98: ...92 8 3 Returning the light source for repair CLV S190 INSTRUCTION MANUAL Ch 8...