App.
104
EMC information
CLV-S190 INSTRUCTION MANUAL
WARNING
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to
any part of this instrument, including cables specified by Olympus. Otherwise,
degradation of the performance of this equipment could result.
Test frequency
[MHz]
Band [MHz]
Modulation
*1
*1 The carrier shall be modulated using a 50% duty cycle square wave signal.
Maximum power
[W]
Immunity test
level [V/m]
385
380 – 390
Pulse modulation
*1
18 Hz
1.8
27
450
430 – 470
FM
5 kHz
deviation 1 kHz sine
2
28
710
704 – 787
Pulse modulation
*1
217 Hz
0.2
9
745
780
810
800 – 960
Pulse modulation
*1
18 Hz
2
28
870
930
1720
1700 – 1990
Pulse modulation
*1
217 Hz
2
28
1845
1970
2450
2400 – 2570
Pulse modulation
*1
217 Hz
2
28
5240
5100 – 5800
Pulse modulation
*1
217 Hz
0.2
9
5500
5785
Summary of Contents for VISERA ELITE CLV-S190
Page 2: ......
Page 6: ...iv Contents CLV S190 INSTRUCTION MANUAL...
Page 20: ...14 Summary of the Functions of the Instrument CLV S190 INSTRUCTION MANUAL...
Page 22: ...16 1 1 Checking the package contents list CLV S190 INSTRUCTION MANUAL Ch 1...
Page 30: ...24 2 3 Rear and side panels CLV S190 INSTRUCTION MANUAL Ch 2...
Page 42: ...36 3 6 Connection to the AC mains power supply CLV S190 INSTRUCTION MANUAL Ch 3...
Page 62: ...56 4 11 After inspection CLV S190 INSTRUCTION MANUAL Ch 4...
Page 98: ...92 8 3 Returning the light source for repair CLV S190 INSTRUCTION MANUAL Ch 8...