44
4.6 Inspection of the examination light
CLV-S190 INSTRUCTION MANUAL
Ch.4
WARNING
Do not look directly into the distal end of the endoscope, the distal end of the light
guide cable, or the output socket of the light source when they are emitting light.
The intense light may cause eye injury.
4.6
Inspection of the examination light
1
Press the lamp button if the examination lamp
is in standby and confirm that the lamp
indicator
“ON” lights up.
Figure 4.5
2
Confirm that the examination light is emitted
from the distal end of the endoscope. When the
lamp light intensity decrease even if the “500 h”
indicator does not light up, replace the
examination lamp with a new one as described
in Section 6.1, “Replacement of the
examination (xenon) lamp”.
Figure 4.6
3
Press and hold the lamp button for about 1 second: the lamp indicator “STBY” lights
up. (See Figure 4.5)
4
Confirm that the examination light is not emitted from the distal end of the endoscope.
Lamp indicator “ON”
Lamp button
Emergency lamp indicator
Distal end of
endoscope
Examination
light
Summary of Contents for VISERA ELITE CLV-S190
Page 2: ......
Page 6: ...iv Contents CLV S190 INSTRUCTION MANUAL...
Page 20: ...14 Summary of the Functions of the Instrument CLV S190 INSTRUCTION MANUAL...
Page 22: ...16 1 1 Checking the package contents list CLV S190 INSTRUCTION MANUAL Ch 1...
Page 30: ...24 2 3 Rear and side panels CLV S190 INSTRUCTION MANUAL Ch 2...
Page 42: ...36 3 6 Connection to the AC mains power supply CLV S190 INSTRUCTION MANUAL Ch 3...
Page 62: ...56 4 11 After inspection CLV S190 INSTRUCTION MANUAL Ch 4...
Page 98: ...92 8 3 Returning the light source for repair CLV S190 INSTRUCTION MANUAL Ch 8...