28
3.3 Installation of equipment
CLV-S190 INSTRUCTION MANUAL
Ch.3
Installation on the mobile workstation (WM-NP2, WM-NP1,
or WM-WP1)
1
Place the mobile workstation on a level surface. Lock the caster brakes by pushing
them down as shown in Figure 3.1.
Figure 3.1
2
According to the combination of equipment, install the shelf of the mobile workstation
as described in the mobile workstation’s instruction manual.
3
Align the two front feet of the light source with
the anti-slip projections on the shelf of the
mobile workstation.
Figure 3.2
4
Peel the paper from the back of the two foot holders of the light source. Place the foot
holders to the two corresponding positions on the rear of the mobile shelf and attach
them lightly. (See Figure 3.2)
5
Remove the light source from the mobile workstation and attach the foot holders
firmly.
6
Place the light source on the mobile shelf so that the rear feet fit into the foot holders.
Caster brakes
Push to lock.
Push to release.
Anti-slip projection
Foot holder
Summary of Contents for VISERA ELITE CLV-S190
Page 2: ......
Page 6: ...iv Contents CLV S190 INSTRUCTION MANUAL...
Page 20: ...14 Summary of the Functions of the Instrument CLV S190 INSTRUCTION MANUAL...
Page 22: ...16 1 1 Checking the package contents list CLV S190 INSTRUCTION MANUAL Ch 1...
Page 30: ...24 2 3 Rear and side panels CLV S190 INSTRUCTION MANUAL Ch 2...
Page 42: ...36 3 6 Connection to the AC mains power supply CLV S190 INSTRUCTION MANUAL Ch 3...
Page 62: ...56 4 11 After inspection CLV S190 INSTRUCTION MANUAL Ch 4...
Page 98: ...92 8 3 Returning the light source for repair CLV S190 INSTRUCTION MANUAL Ch 8...