6.1 Replacement of the examination (xenon) lamp
CLV-S190 INSTRUCTION MANUAL
Ch.6
CAUTION
• Do not touch the glass surface (filter) inside of the lamp chamber. Natural skin
moisture from your fingers can cause cracks and damage the light source.
• Do not touch the glass surface of the lamp, mesh, or reflector. Natural skin moisture
from your fingers can cause cracks and damage the light source. If the glass
surface of the lamp gets dirty, wipe it off with a lint-free cloth.
• Handle the lamp carefully. Do not apply excessive force or scratch the examination
lamp. The glass surface may crack, the lamp life may shorten, or the light source
may be damaged due to the high pressure inside the lamp.
• When replacing the examination lamp, use a clean, lint-free cloth to wipe off
residual heat compound from the heat sink. If the heat compound is not wiped off
completely, the lamp’s heat efficiency will be impaired and the examination lamp life
will be shortened significantly.
• Be sure to reset the lamp usage indicator as described in Section 6.4, “Lamp usage
indicator reset” after replacing.
Otherwise, the incorrect total working hours of the examination lamp will be
indicated.
Summary of Contents for VISERA ELITE CLV-S190
Page 2: ......
Page 6: ...iv Contents CLV S190 INSTRUCTION MANUAL...
Page 20: ...14 Summary of the Functions of the Instrument CLV S190 INSTRUCTION MANUAL...
Page 22: ...16 1 1 Checking the package contents list CLV S190 INSTRUCTION MANUAL Ch 1...
Page 30: ...24 2 3 Rear and side panels CLV S190 INSTRUCTION MANUAL Ch 2...
Page 42: ...36 3 6 Connection to the AC mains power supply CLV S190 INSTRUCTION MANUAL Ch 3...
Page 62: ...56 4 11 After inspection CLV S190 INSTRUCTION MANUAL Ch 4...
Page 98: ...92 8 3 Returning the light source for repair CLV S190 INSTRUCTION MANUAL Ch 8...