6.2 Removal of the lamp
73
CLV-S190 INSTRUCTION MANUAL
Ch.6
6.2
Removal of the lamp
1
When the examination lamp is ON, press and hold the lamp button for about 1 second
to extinguish it.
2
Wait for a few minutes to allow the light source to cool the lamp chamber sufficiently.
NOTE
While the light source is ON and the examination lamp is extinguished, a fan built
into the light source cools the lamp chamber.
3
Turn the light source OFF.
4
Disconnect the power cord from the hospital-grade wall mains outlet.
5
Turn the knobs of the lamp cover and remove the lamp cover.
Figure 6.1
6
Confirm that the inside of the lamp chamber is not too hot. If the lamp chamber is
extremely hot, attach the lamp cover, connect the power cord, turn the light source
ON, and then repeat Step 1 through 6 above.
Knob
Lamp cover
Lamp cover