4.4 Inspection of the power supply
41
CLV-S190 INSTRUCTION MANUAL
Ch.4
Confirm that the ventilation grills are not covered with dust, that the lamp cover is attached, and that
the light source is turned ON.
WARNING
Turning ON the light source ignites the examination lamp automatically. Do not look
directly into the distal end of the endoscope or the output socket of the light source
when they are emitting light. Eye injury may result.
4.4
Inspection of the power supply
1
Confirm that the ventilation grills on the side and rear panels of the light source are
not covered with dust or other materials.
2
Confirm that the lamp cover is attached firmly.
Figure 4.3
NOTE
For safety reasons, incorrect attachment of the lamp cover prevents the light
source from being turned ON.
3
Press the power switch of the light source. (See Figure 4.3)
4
Confirm that the power indicator lights up. (See Figure 4.3)
5
Confirm that the emergency lamp inductor on the control panel is not lighting or
blinking. (See Figure 4.3)
6
Confirm that air is exhausted by holding your hand in front of the ventilation grills on
the rear panel.
Control panel
Power indicator
Power switch
Lamp cover
Summary of Contents for VISERA ELITE CLV-S190
Page 2: ......
Page 6: ...iv Contents CLV S190 INSTRUCTION MANUAL...
Page 20: ...14 Summary of the Functions of the Instrument CLV S190 INSTRUCTION MANUAL...
Page 22: ...16 1 1 Checking the package contents list CLV S190 INSTRUCTION MANUAL Ch 1...
Page 30: ...24 2 3 Rear and side panels CLV S190 INSTRUCTION MANUAL Ch 2...
Page 42: ...36 3 6 Connection to the AC mains power supply CLV S190 INSTRUCTION MANUAL Ch 3...
Page 62: ...56 4 11 After inspection CLV S190 INSTRUCTION MANUAL Ch 4...
Page 98: ...92 8 3 Returning the light source for repair CLV S190 INSTRUCTION MANUAL Ch 8...