Important Information — Please Read Before Use
11
CLV-S190 INSTRUCTION MANUAL
Cardiac applications
DANGER
• Use only the devices listed in the “
System chart” on page 93 for endoscopic
observation or treatment of the heart or areas near the heart. Other combinations
of equipment may cause ventricular fibrillation or seriously affect the cardiac
function of the patient.
• For cardiac applications, never support the endoscope with a metal surgical arm
that is not electrically isolated from the ground. If not isolated, the endoscope will
be connected to the ground through the surgical arm and bed, and will conduct
unexpected leakage current that may seriously affect the cardiac function of the
patient.
• The use of medical devices not specifically designed for cardiac applications may
cause ventricular fibrillation or seriously affect the cardiac function of the patient. As
specified by the international standard IEC 60601-1, any applied part used for
observation or treatment of the heart or areas near the heart must meet “TYPE CF
applied part” requirements for low electrical leakage current. When using
endoscopes for endoscopic cardiac applications, the applied part requirements
include all devices directly connected to the endoscope, such as the light guide
cable, camera head, and telescope holder. Each of these devices must individually
meet the “TYPE CF applied part” requirements for leakage current limits if they are
to be used for cardiac applications.
NOTE
• The OLYMPUS light guide cables and camera heads listed in the “
System chart”
on page 93 (TYPE CF applied part) that are suitable for cardiac applications bear a
mark.
• The Olympus Surgical Holder for Telescope (SH-1) has an electrically isolated arm
structure that isolates the endoscope from the ground. This design makes the SH-1
suitable for cardiac applications.
Summary of Contents for VISERA ELITE CLV-S190
Page 2: ......
Page 6: ...iv Contents CLV S190 INSTRUCTION MANUAL...
Page 20: ...14 Summary of the Functions of the Instrument CLV S190 INSTRUCTION MANUAL...
Page 22: ...16 1 1 Checking the package contents list CLV S190 INSTRUCTION MANUAL Ch 1...
Page 30: ...24 2 3 Rear and side panels CLV S190 INSTRUCTION MANUAL Ch 2...
Page 42: ...36 3 6 Connection to the AC mains power supply CLV S190 INSTRUCTION MANUAL Ch 3...
Page 62: ...56 4 11 After inspection CLV S190 INSTRUCTION MANUAL Ch 4...
Page 98: ...92 8 3 Returning the light source for repair CLV S190 INSTRUCTION MANUAL Ch 8...