4.9 Inspection of the optical-digital observation function

51

CLV-S190  INSTRUCTION MANUAL

Ch.4

Optical-digital observation is available when the light source is used in combination with the 
videoscope or camera head compatible with the optical-digital observation. Confirm that the 
optical-digital observation mode is switched.
The PDD observation is available when installing an optional PDD filter (MAJ-1429) in the light source.

4.9

Inspection of the optical-digital observation 
function

1

Confirm that the observation mode selection 
indicators show the available optical-digital 
observation modes.

Figure 4.15

NOTE

The available observation modes vary depending on the connected endoscope.

2

Press the observation mode button on the front 
panel to light up the observation mode indicator 
“ON”. Confirm that the endoscopic image is 
displayed in the optical-digital observation 
indicated by the observation mode selection 
indicator.

Figure 4.16

3

Press the observation mode button. The observation mode indicator “STBY” lights up, 
and the instrument returns to the normal light observation mode. (See Figure 4.16)

Observation mode 
selection indicators

Observation mode 
indicators

Observation mode button

Observation mode 
selection indicators

Summary of Contents for VISERA ELITE CLV-S190

Page 1: ...4 Summary of the Functions of the Instrument 12 Chapter 1 Checking the Package Contents 15 Chapter 2 Nomenclature and Functions 17 Chapter 3 Installation and Connection 25 Chapter 4 Inspection 37 Chap...

Page 2: ......

Page 3: ...ackage Contents 15 1 1 Checking the package contents list 15 Chapter 2 Nomenclature and Functions 17 2 1 Nomenclature and functions 17 2 2 Front panel 19 2 3 Rear and side panels 22 Chapter 3 Installa...

Page 4: ...e optical digital observation function 51 4 10 Inspection of the high intensity mode 53 4 11 After inspection 55 Chapter 5 Operation 57 5 1 Attention to operation 57 5 2 Operation flow 60 5 3 Turning...

Page 5: ...nation lamp 86 The light source or any of its components 86 Chapter 8 Troubleshooting 87 8 1 Troubleshooting 87 8 2 Troubleshooting guide 87 8 3 Returning the light source for repair 91 Appendix 93 Co...

Page 6: ...iv Contents CLV S190 INSTRUCTION MANUAL...

Page 7: ...mbols are missing or illegible contact Olympus Front side Labels and Symbols OPT DGTL MODE LAMP ON STBY NBI PDD HIGH ON STBY BRIGHTNESS AUTO MANU MIN MAX 0 SPARE LAMP COUNTER RESET 0 100h 300h 500h UD...

Page 8: ...bol Indicates to see the instruction manual for the cable to be connected Cautions that the inside is extremely hot immediately after the lamp is extinguished Potential equalization terminal Product n...

Page 9: ...Labels and Symbols 3 CLV S190 INSTRUCTION MANUAL Back cover of this instruction manual Symbol Description Manufacturer Authorized representative in the European Community...

Page 10: ...s are greater than its risks Fully explain to the patient the potential benefits and risks of the endoscopy and endoscopic treatment as well as any examination treatment methods that can be performed...

Page 11: ...high leakage currents to decrease the possibility of electric shock Automatic brightness adjustment The automatic brightness adjustment automatically adjusts the intensity of the light emitted from t...

Page 12: ...s Instrument compatibility Refer to the System chart on page 93 to confirm that the light source is compatible with the ancillary equipment being used Using incompatible equipment can result in patien...

Page 13: ...the light source stop operation of the light source immediately and contact Olympus Do not prepare inspect or use the light source with wet hands Never install and operate the light source in the fol...

Page 14: ...um required brightness The brightness of the image on a video monitor may differ from the actual brightness at the distal end of an endoscope Especially in combination with endoscopes using an electri...

Page 15: ...output connector of the light source Operator or patient injury can result Do not connect any object other than the light guide and endoscope to the output connector Otherwise malfunction may result...

Page 16: ...urce Do not position a medical device so that it is difficult to disconnect it from the wall mains outlet NOTE As defined by the international safety standard IEC 60601 1 medical electrical equipment...

Page 17: ...or seriously affect the cardiac function of the patient As specified by the international standard IEC 60601 1 any applied part used for observation or treatment of the heart or areas near the heart...

Page 18: ...he light source is used in combination with a fiberscope the examination light intensity has to be adjusted manually Section 5 5 Brightness adjustment Optical digital observation NBI observation is av...

Page 19: ...mergency lamp If the examination lamp does not light up or blows in the middle of an examination making endoscopic observation impossible the light source switches automatically to the emergency light...

Page 20: ...14 Summary of the Functions of the Instrument CLV S190 INSTRUCTION MANUAL...

Page 21: ...wn below Inspect each item for damage If the light source is damaged a component is missing or you have any questions do not use the light source and immediately contact Olympus 1 1 Checking the packa...

Page 22: ...16 1 1 Checking the package contents list CLV S190 INSTRUCTION MANUAL Ch 1...

Page 23: ...2 1 Nomenclature and functions Symbol Description Symbol Description Power ON OFF Brightness Decrease brightness Increase brightness Exchange lamp lamp usage indicator Lamp usage indicator reset New...

Page 24: ...LV S190 INSTRUCTION MANUAL Ch 2 Rear and side panels Symbol Description Symbol Description Refer to instructions Warns that the lamp is hot Serial number Potential equalization terminal Alternating cu...

Page 25: ...ator Lights up when the light source is ON 3 Output socket Connect the endoscope the endoscope s light connector or the light guide cable to this socket This socket provides light to the endoscope 4 C...

Page 26: ...el CLV S190 INSTRUCTION MANUAL Ch 2 Control panel buttons OPT DGTL MODE LAMP ON STBY NBI PDD HIGH ON STBY BRIGHTNESS AUTO MANU MIN MAX 0 SPARE LAMP COUNTER RESET 0 100h 300h 500h 5 6 7 8 9 10 11 12 13...

Page 27: ...and igniting the examination lamp Section 5 8 Extinguishing the examination lamp 11 Lamp indicators These indicators display whether the examination lamp xenon lamp lights up ON or not STBY Section 5...

Page 28: ...22 2 3 Rear and side panels CLV S190 INSTRUCTION MANUAL Ch 2 Rear panel Side panel 2 3 Rear and side panels 1 6 2 4 3 5 8 7...

Page 29: ...tion of the foot switch 4 Potential equalization terminal This terminal is connected to a potential equalization terminal of the other equipment connected to the light source The electrical potential...

Page 30: ...24 2 3 Rear and side panels CLV S190 INSTRUCTION MANUAL Ch 2...

Page 31: ...F all system components before connecting them Otherwise equipment damage or malfunction can result Use appropriate cables only Otherwise equipment damage or malfunction can result Properly and secure...

Page 32: ...on workflow 1 Install the light source and the ancillary equipment on the mobile workstation or another stable location Section 3 3 on page 27 2 Connect the light source to the video system center Onl...

Page 33: ...ot holders MAJ 1205 Otherwise the light source may topple down or drop and operator or patient injury may occur or equipment damage can result When using the mobile workstation confirm that excessive...

Page 34: ...manual 3 Align the two front feet of the light source with the anti slip projections on the shelf of the mobile workstation Figure 3 2 4 Peel the paper from the back of the two foot holders of the li...

Page 35: ...tion location Peel the paper from the bottom of the foot holders and lightly attach the foot holders to the sheet Figure 3 3 2 Place the light source on the pattern sheet and check that the feet fit i...

Page 36: ...Ch 3 Compatible video system center Table 3 1 3 4 Connection of the video system center Connect the video system center to the light source using the cables listed below Model Product name OTV S190 V...

Page 37: ...the video system center 31 CLV S190 INSTRUCTION MANUAL Ch 3 OTV S190 Cable Table 3 2 Connecting diagram Figure 3 4 Model Product name MAJ 1959 LIGHT SOURCE CABLE OTV terminal Light source terminal OTV...

Page 38: ...f the video system center CLV S190 INSTRUCTION MANUAL Ch 3 OTV S7Pro Cable Table 3 3 Connecting diagram Figure 3 5 Model Product name MAJ 1411 LIGHT SOURCE CABLE OTV terminal Light source terminal OTV...

Page 39: ...ervation is not intended it is recommended to connect the foot switch to the video system center instead of the light source If naked eye PDD observation is intended connect the foot switch to the lig...

Page 40: ...ty is insufficient a fire can result or the circuit breaker may trip and it may turn OFF not only the light source but also all other equipment connected to the same power circuit When using the mobil...

Page 41: ...t may vary from this drawing as in each country the commercialized cable is in accordance to the local applicable norm 1 Confirm that the power switch is not pressed in 2 Connect the power cord to the...

Page 42: ...36 3 6 Connection to the AC mains power supply CLV S190 INSTRUCTION MANUAL Ch 3...

Page 43: ...ved do not use the light source and see Chapter 8 Troubleshooting If the light source with any irregularity is used the light source may malfunction or be damaged and an electric shock and or burns ca...

Page 44: ...operating hours of the examination lamp Section 4 5 on page 43 4 Confirm that the examination lamp turns ON OFF normally Section 4 6 on page 44 5 Check the brightness mode Section 4 7 on page 45 6 In...

Page 45: ...e light guide cable or the distal end of the endoscope becomes very hot To prevent a fire hazard do not bring the disconnected end of the light guide cable or distal end of the endoscope in contact wi...

Page 46: ...de cable to the rigid endoscope 3 Insert the light guide connector into the output socket on the front panel of the light source until it clicks Figure 4 1 1 Inspect the endoscope as described in the...

Page 47: ...that the ventilation grills on the side and rear panels of the light source are not covered with dust or other materials 2 Confirm that the lamp cover is attached firmly Figure 4 3 NOTE For safety rea...

Page 48: ...n 5 seconds automatically If ignition fails the light source automatically switches to the emergency lamp and the emergency lamp indicator lights up CAUTION If the emergency lamp instead of the examin...

Page 49: ...f the examination xenon lamp e g 500 h means 500 hours 4 5 Checking the lamp usage indicator Check the lamp usage indicator When the 500 h indicator on the lamp usage indicator lights up replace the e...

Page 50: ...p indicator ON lights up Figure 4 5 2 Confirm that the examination light is emitted from the distal end of the endoscope When the lamp light intensity decrease even if the 500 h indicator does not lig...

Page 51: ...mode to MANU Setting it to AUTO does not enable the automatic brightness adjustment and the brightness may not be adequate 4 7 Inspection of the brightness mode switching function 1 Confirm that the...

Page 52: ...as shown in Table 4 1 Brightness adjustment mode Connected endoscopes Reference AUTO Rigid videoscope Flexible videoscope Rigid endoscope used in combination with the camera head and the light guide c...

Page 53: ...head is disconnected while the brightness mode indicator is set to AUTO the intense light may cause eye injury 1 Confirm that the endoscope is connected to the light source and the video system center...

Page 54: ...n 5 and 60 mm Confirm that the light being emitted from the distal end varies with the distance Figure 4 10 5 Hold the distal end of the endoscope at a distance between 30 and 40 mm to the object and...

Page 55: ...s adjustment always set the brightness to the minimum level necessary to complete the examination If the light is too bright eye injury or burns can result 1 Press the lamp button on the control panel...

Page 56: ...on the light source varies in an interlocked operation 3 Press the brightness buttons or and confirm the following Each time either of the brightness buttons is pressed a beep is heard and the bright...

Page 57: ...ion mode selection indicators show the available optical digital observation modes Figure 4 15 NOTE The available observation modes vary depending on the connected endoscope 2 Press the observation mo...

Page 58: ...t Olympus for details on installing the optional PDD filter MAJ 1429 When the foot switch MAJ 1391 optional is connected to the light source the observation mode can be switched during naked eye PDD o...

Page 59: ...he MANU indicator lights up Figure 4 17 2 Press the brightness buttons on the control panel to set the brightness level to 0 Figure 4 18 3 Press the intensity mode button on the control panel the norm...

Page 60: ...sity setting is automatically saved when the light source is turned OFF and is recalled when the light source is turned ON again 4 Confirm that the high intensity mode indicator lights up and that the...

Page 61: ...ton on the control panel for about 1 second to turn OFF the examination lamp refer to Section 5 8 Extinguishing the examination lamp Figure 4 20 2 If the light source will not be used immediately pres...

Page 62: ...56 4 11 After inspection CLV S190 INSTRUCTION MANUAL Ch 4...

Page 63: ...on of the light source Before using the light source make sure to perform inspection of the light source as described in Chapter 4 Inspection WARNING Be sure to wear protective equipment such as eye w...

Page 64: ...ect the endoscope to a spare light source Note that the emergency lamp offers only the minimum brightness required in case of an emergency and it is dangerous to continue the use of the light source u...

Page 65: ...the observation mode For PDD observation use only the light guide cable and rigid endoscope compatible with PDD Otherwise an improper image may be obtained NOTE Set the brightness of the examination...

Page 66: ...brightness mode Section 5 4 on page 62 4 Adjust the brightness of the examination lamp Section 5 5 on page 63 5 Operate the light source Optical digital observation Section 5 6 on page 66 High intens...

Page 67: ...sult NOTE When the video system center is connected either the light source or the video system center can be turned ON first 5 3 Turning the light source ON and igniting the examination lamp 1 Confir...

Page 68: ...adequate 5 4 Setting the brightness mode 1 Confirm the brightness adjustment mode of the endoscope to be used according to the Table 5 1 2 Press the brightness mode button to select the brightness mod...

Page 69: ...or to the minimum level If the camera head is disconnected while the brightness mode is set to AUTO the intense light may injure your eyes CAUTION Turn the video system center ON to enable the automat...

Page 70: ...once moves the level one step pressing the button continuously moves the level continuously The brightness level value is saved per observation mode The brightness level should be adjusted for each ob...

Page 71: ...period Use of higher brightness than required may cause eye injury or burns to the patient NOTE Pressing either of the brightness buttons or once increases or decreases the level by one step pressing...

Page 72: ...f lesions or for a decision regarding any potential diagnostic or therapeutic intervention If the endoscopic image seems to be dark in the NBI observation change to the normal observation Otherwise an...

Page 73: ...gs of the video system center The settings for observation mode are not retained after the light source is turned OFF The light source is always in the normal light observation mode at the moment it i...

Page 74: ...ble when no endoscope is connected to the light source The function is available only when a compatible endoscope is connected Refer to Section 4 10 Inspection of the high intensity mode for the compa...

Page 75: ...lamp 69 CLV S190 INSTRUCTION MANUAL Ch 5 5 8 Extinguishing the examination lamp Press and hold the lamp button for about 1 second the examination lamp turns OFF and the indicator STBY lights up Figure...

Page 76: ...end of the light guide cable or distal end of the endoscope in contact with a flammable object such as operating room drapes when the examination lamp is turned ON When no examination is performed be...

Page 77: ...lamp chamber is extremely hot immediately after the lamp is turned OFF Operator and or patient burns can result When replacing the lamp do not leave any objects such as a cloth and plastic bag inside...

Page 78: ...amp carefully Do not apply excessive force or scratch the examination lamp The glass surface may crack the lamp life may shorten or the light source may be damaged due to the high pressure inside the...

Page 79: ...ON and the examination lamp is extinguished a fan built into the light source cools the lamp chamber 3 Turn the light source OFF 4 Disconnect the power cord from the hospital grade wall mains outlet 5...

Page 80: ...Ch 6 7 Remove the hexagon wrench from the rear of the lamp cover Figure 6 2 8 Turn knob A counterclockwise by 90 to loosen it Figure 6 3 9 Turn knob B counterclockwise by 90 to loosen it See Figure 6...

Page 81: ...amp with heat sinks A and B attached Figure 6 4 11 Using the hexagon wrench loosen the three bolts on heat sink B on the side of the examination lamp and remove heat sink B from the examination lamp F...

Page 82: ...e lamp s heat efficiency will be impaired and the examination lamp life will be shortened significantly CAUTION Washers are attached to bolts Retain all three of them carefully because they will be us...

Page 83: ...t enough heat compound is applied the heat can cause lamp ignition failures 6 3 Insertion of the lamp 1 Hold the new examination lamp without touching the glass surface 2 Using your finger apply the h...

Page 84: ...the life of the examination lamp Confirm the direction of the washers when tightening the bolts If a washer is positioned incorrectly the lamp may fail to light properly Figure 6 9 4 Using your finger...

Page 85: ...tighten the heat sink clamp firmly Poor heat radiation may result in equipment damage examination lamp ignition failure and a considerable drop in the life of the examination lamp 7 Insert both heat...

Page 86: ...e horizontal position until it stops while pushing the knob See Figure 6 13 10 Confirm that the heat sinks are attached firmly by pulling the projection See Figure 6 12 CAUTION If the heat sinks are n...

Page 87: ...hen the examination lamp is not exchanged The incorrect working hours of the examination lamp will be indicated 12 Turn the knobs of the lamp cover and close the lamp cover securely Figure 6 15 NOTE F...

Page 88: ...tton on the control panel for more than 3 seconds the lamp usage indicator is reset Confirm that the lamp usage indicator shows 0 Figure 6 17 NOTE The counter reset button is inactive while the examin...

Page 89: ...ly spray type medical agents such as rubbing alcohol directly to the light source Medical agents might enter the light source through the ventilation grills and may cause equipment damage CAUTION Do n...

Page 90: ...each country the commercialized cable is in accordance to the local applicable norm 3 If the equipment is soiled with blood or other potentially infectious materials wipe off all gross debris using a...

Page 91: ...adio equipment or cellular portable phone Damage to the light source may result 7 2 Storage 1 Confirm that the light source is OFF 2 Confirm that the power cord is disconnected from the hospital grade...

Page 92: ...uld burst and scatter in all directions resulting in injury The protective equipment includes eye wear face mask and gloves The light source or any of its components Dispose of the light source or any...

Page 93: ...rce and contact Olympus WARNING Never use the light source if an irregularity is observed Damage or irregularity of the light source may result equipment damage an electric shock and or burns NOTE If...

Page 94: ...mp The examination lamp is broken Replace the examination lamp with a new one as described in Section 6 1 Replacement of the examination xenon lamp No light is emitted from the endoscope The examinati...

Page 95: ...5 5 Brightness adjustment or turn the light source OFF then ON again The field of view and the image are too dark or too bright The examination lamp is old Replace the examination lamp with a new one...

Page 96: ...ndoscope light guide cable and camera head compatible with PDD The video system center does not have the PDD observation function Use the video system center with the PDD observation function The PDD...

Page 97: ...the individual at your location who is the most familiar with the problem Also include a repair purchase order CAUTION Olympus is not liable for any injury or damage that occurs as a result of repair...

Page 98: ...92 8 3 Returning the light source for repair CLV S190 INSTRUCTION MANUAL Ch 8...

Page 99: ...ight source are listed below New products released after the introduction of the light source may also be compatible for use in combination with the light source For further details contact Olympus WA...

Page 100: ...es Flexible videoscope ENF series CYF series HYF series LTF series LF series Camera head 1 OTV S7H series Light guide cable 1 Camera head 1 CH S190 series OTV S7ProH series OTV S7H series 1 The OLYMPU...

Page 101: ...on WM NP2 WM NP1 WM WP1 VISERA ELITE video system center OTV S190 VISERA Pro video system center OTV S7Pro Keyboard VISERA ELITE xenon light source CLV S190 High definition LCD monitor OEV261H OEV191H...

Page 102: ...ature 5 40 C 41 104 F Relative humidity 10 90 without condensation Atmospheric Pressure 700 1060 hPa 0 7 1 1 kgf cm2 10 2 15 4 psia Transportation and storage environment Ambient temperature 25 to 70...

Page 103: ...y 500 hours Automatic brightness adjustment Automatic brightness adjustment method Servo diaphragm method Automatic exposure 17 steps Indicators on front panel Emergency lamp Indicates absence of emer...

Page 104: ...ies with the EMC standard for medical electrical equipment edition 4 IEC 60601 1 2 2014 When connecting to an instrument that complies with a previous edition of the EMC standard for medical electrica...

Page 105: ...y for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment Radiated emissions CISPR 11 Class B This instrument s RF...

Page 106: ...D IEC 61000 4 2 Contact 8 kV Air 2 4 8 15 kV Contact 2 4 6 kV Air 2 4 8 kV Same as left Floors should be made of wood concrete or ceramic tile that hardly produces static If floors are covered with sy...

Page 107: ...a battery 40 UT 60 dip in UT for 5 cycle 70 UT 30 dip in UT for 25 cycle 50 Hz 30 cycle 60 Hz Phase angle causing voltage dips 0 70 UT 30 dip in UT for 25 cycle 0 UT 100 dip in UT for 250 cycle 50 Hz...

Page 108: ...ed separation distance Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m 6V IS...

Page 109: ...st and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the me...

Page 110: ...quipment could result Test frequency MHz Band MHz Modulation 1 1 The carrier shall be modulated using a 50 duty cycle square wave signal Maximum power W Immunity test level V m 385 380 390 Pulse modul...

Page 111: ...reproduced or distributed without the express written permission of OLYMPUS MEDICAL SYSTEMS CORP OLYMPUS is a registered trademark of OLYMPUS CORPORATION Trademarks product names logos or trade names...

Page 112: ...Valley Road 12 01 04 Valley Point Office Tower Singapore 248373 Fax 6834 2438 Telephone 6834 0010 A8F Ping An International Financial Center No 1 3 Xinyuan South Road Chaoyang District Beijing 100027...

Reviews: