38
4.2 Inspection workflow
CLV-S190 INSTRUCTION MANUAL
Ch.4
Follow each step of the workflow for inspection of the light source before use.
4.2
Inspection workflow
1
Connect an endoscope.
Section 4.3 on page 39
2
Confirm that the light source is turned ON normally.
Section 4.4 on page 41
3
Check the total accumulated operating hours of the examination
lamp.
Section 4.5 on page 43
4
Confirm that the examination lamp turns ON/OFF normally.
Section 4.6 on page 44
5
Check the brightness mode.
Section 4.7 on page 45
6
Inspect the necessary functions (dependent upon the
endoscope).
• Brightness level
Section 4.8 on page 46
• Optical-digital observation
Section 4.9 on page 51
• High intensity mode
Section 4.10 on page 53
7
After inspection
Turn the examination lamp OFF if the light source will not be
used immediately.
Section 4.11 on page 55