6.1 Replacement of the examination (xenon) lamp
71
CLV-S190 INSTRUCTION MANUAL
Ch.6
Chapter 6
Lamp Replacement
Always use the examination lamp designated below. To order a new examination lamp, contact
Olympus.
• Xenon lamp MD-631
WARNING
• Never install a lamp that has not been approved by Olympus. The use of a
nonapproved lamp can cause damage to the light source and ancillary equipment,
malfunction or a fire.
• Turn the light source OFF and disconnect the power cord from the wall mains outlet
before replacing the lamp with a new one. Otherwise, an electric shock may result.
• Do not touch anything inside the lamp chamber. The lamp chamber is extremely
hot immediately after the lamp is turned OFF. Operator and/or patient burns can
result.
• When replacing the lamp, do not leave any objects (such as a cloth and plastic bag)
inside the lamp chamber. A fire and/or equipment damage can result.
• Store the hexagon wrench securely on the rear side of the lamp cover. If the
wrench drops or any objects get inside the light source, turn OFF the light source
immediately, disconnect the power cord and contact Olympus. If the light source is
used while the wrench is left inside the light source, equipment damage and/or an
electric shock can result.
• Dispose of the examination lamp as described in Section 7.3, “Disposal”. The glass
surface may crack due to the high pressure inside the lamp if disposed of
improperly.
6.1
Replacement of the examination (xenon) lamp
Summary of Contents for VISERA ELITE CLV-S190
Page 2: ......
Page 6: ...iv Contents CLV S190 INSTRUCTION MANUAL...
Page 20: ...14 Summary of the Functions of the Instrument CLV S190 INSTRUCTION MANUAL...
Page 22: ...16 1 1 Checking the package contents list CLV S190 INSTRUCTION MANUAL Ch 1...
Page 30: ...24 2 3 Rear and side panels CLV S190 INSTRUCTION MANUAL Ch 2...
Page 42: ...36 3 6 Connection to the AC mains power supply CLV S190 INSTRUCTION MANUAL Ch 3...
Page 62: ...56 4 11 After inspection CLV S190 INSTRUCTION MANUAL Ch 4...
Page 98: ...92 8 3 Returning the light source for repair CLV S190 INSTRUCTION MANUAL Ch 8...