CURR:UCL {<numeric value>}
This command sets the under current limit level.
0V ~ currently setting value.(up to current setting value).
ex) curr 10
At first set the current value
curr:ucl 5
Set the lower value than above setting value
From the time set UCL, you can not set the current to below 5A.
CURR:UCL?
This query returns the under current limit value presently programmed.
Return value
"numeric value"
ex) curr:ucl?
return value "5.00"
CURR:OCL {<numeric value>}
This command sets the over current limit level.
> numeric value
Currently setting value~miximum of limit value.
ex) curr 10
At first set the current value
curr:ocl 15
Set the higher value than above setting value
From the time set OCL, you can not set the current to above 15A.
CURR:OCL?
This query returns the overr current limit value presently programmed.
Return value
"numeric value"
ex) curr:ocl?
return value "15.00"
> numeric value