-23, "En route to cal the curr"
It appears when sended calibration command about voltage during current calibration.
-24, "Over volt min parameter"
It appears voltage Min value is over the power supply's voltage limit.
Refer to "4-5. Using Front Panel Calibration."
-25, "Under volt max parameter"
It appears voltage Max value is gotton out the power supply's voltage lowest limit value.
Refer to "4-5. Using Front Panel Calibration."
-26, "Over volt max parameter"
It appears voltage Max value is gotton out the power supply's voltage highest limit value.
Refer to "4-5. Using Front Panel Calibration." .
-27, "Ignored min run under curr"
If you set the "max" command or value input of calibration without current "Min".
Procedure : Min → VALUE → MAX → VALUE
-28, "Ignored min save under curr"
If you set the "max" command without "Value" of current "Min" calibration operating.
Procedure : Min → VALUE → MAX → VALUE
-29, "Invalid min value use under curr"
After sent the command of current "Min" value and then you sent the value command without
"Max" command calibration operating.
Procedure : Min → VALUE → MAX → VALUE
-30, "En route to cal the curr"
It appears when sended calibration command about currrent during voltage calibration.
-31, "Over curr min parameter"
It appears current Min value is over the power supply's current limit.
Refer to "4-5. Using Front Panel Calibration." .
-32, "Under curr max parameter"
It appears currenr Max value is gotton out the power supply's current lowest limit value.
Refer to "4-5. Using Front Panel Calibration." .
-33, "Over curr max parameter"
It appears current Max value is gotton out the power supply's current highest limit value.
Refer to "4-5. Using Front Panel Calibration." .
-34, "Not allowed command under cal"
Under the remote calibration mode, you can not use another command.