4-3. Equipment required for Calibration
We recommend as below measuring instruments in other to calibrate the power supply.
4-4. Technic of Calibratio
As below is for calibration technics.
Instruments Connection Drawing.
▌
Connect as Figure. 4-1.
▌
When calibrates voltage, turn off the electronic load or remove the connection from output bus bar.
▌
It is fine to use the general load resistor instead of electronic load.
< Figure 4-1 >
Using article
0.001Ω , 0.01%
Measuring ripple & noise
Current
monitoring
- 55 -
Oscilloscope
Agilent 34401A
Resolution: 0.1 mV
Accuracy: 0.01%
Requested performance
Recommended
Voltage Range: 500 Vdc
Current Range: 120 Adc
Open and Short Switches
Transient On/Off
Current calibration
for Power Supply
(General load resistor is also
OK)
M/I
Tektronix
TDS3014
Electronic Load
When current calibration,
100 MHz with
20MHz bandwidth
Digital
voltmeter
Voltage Calibration