Calculate the current when DVM display voltage became stable.
If resistor is 0.01Ω and DVM display voltage value is 1.23mV, the current
value is 0.123A.
▌
Enters a current value that you calculated by using cursor key and
encoder knob.
LCD Display
a-LOW 00.12a
This is a example of 60V
▌
After finishing to enter the value, press the "Calibration" key.
LCD Display
ADC datA 026ch
About 500ms, display HEX value instead
of readback value.
LCD Display
3.cal-curr low
Return to the previous state of calibration
▌
Move to "4.cal-curr High" mode by using cursor key and press the "Calibration" key
LCD Display
a-high 61.00a
Display current can be different
Wait for 5 minute
▌
Wait until DVM display voltage becomes stable
▌
Calculate the current when DVM display voltage became stable
If resistor is 0.01Ω and DVM display voltage value is 615mV, the current
value is 61.50A.
▌
Enters a current value calculated by using cursor key and encoder knob.
LCD Display
a-high 61.50a
This is a example of 60V
▌
After finished to enter the value, press the "Calibration" key.
LCD Display
ADC datA f500h
About 500ms, display HEX value instead
of readback value.
LCD Display
4.cal-curr high
Return to the previous state of calibration
▌
Current caliration is finished. It can be applied next operation of power supply.