7. Error Messages
About error explanation. You can confirm the error by using front panel's "ERROR I/O Local" key
or PC remote nterface.(Refer to "SYST:ERR?" command.)
+0,"No error"
No error state.
7-1. Operating Error
-10, "Invalid the DAC parameter"
In case of out of DAC memory section, it is different between setting value and real output
value and at that time , you must remove the DUT.
It is caused from wrong calibration data. Please operate calibration one more.
Refer to "4. CALIBRATION" .
7-2. Hardware Error
-200, "System interface error"
It appears when SCPI Module is out of order.
-201, "ADC operating failed"
It appears when circuit of ADC part is out of order.
-202, "Front panel operating failed"
It appears when front panel does not respose.
-255, "Error not define"
It appears when unknown error occurs.
7-3. Remote Calibration Error
Refer to "4-6. Remote Interface Calibration."
-20, "Ignored min run under volt"
If you set the "max" command or value input of calibration without voltage "Min"
calibration operating at first.
Procedure : Min → VALUE → MAX → VALUE
-21, "Ignored min save under volt"
If you set the "max" command without "Value" of voltage "Min" calibration operating.
Procedure : Min → VALUE → MAX → VALUE
-22, "Invalid min value use under volt"
After sent the command of voltage "Min" value and then you sent the value command without
"Max" command calibration operating.
Procedure : Min → VALUE → MAX → VALUE