1
.
Inspection
32
No.704-2010-01
③
Mounting the arm
In a state in which the arm is pushed,
check that the tray moves smoothly.
Adjust the mounting position of the linear
guide which the arm to come to the center
of the case.
④
Table position
Make sure that the protrusion of the arm is
housed in the hole in the tray when the tray
is raised.
It is to be performed after the adjustment all
finished.
Open the Parameter screen. And input
“0219”.
Adjust at “OFFSET-X(TABLE)