1
.
Inspection
15
No.704-2010-01
2.4 Replacement procedure of H3 heater
(1) Preparation of tools/parts
Turn the main power OFF. Do not start working when the temperature of each heater
is high. (Start working after 2 hours from turning the power OFF when the temperature
is high.)
Pull up the ring latch and remove the top cover panel.
Pull out the sample heating tube and the mercury collection tube from the front.
Remove the cover of H3 terminal board. (Two M3 screws)
No.
Name
Part No.
Q’ty
Remarks
①
Philips screw driver (02)
1
Tool
②
Long-nose plier
1
Tool
③
Ball-point screw driver
1
Accessory
④
Adjustable wrench (small)
1
Tool
Terminal board cover