1
.
Inspection
5
No.704-2010-01
1.2.
Cell cleaning
(1) Preparation of components
(2) Removal
Remove the top cover and remove the cover of the optical system from its case. Then,
remove the cell retainers, loosen the joint and the screws fixing the optical system,
and remove the cell unit.
MA-3000 (Version I type)
MA-3000 (Version II type)
* If the 1 mm cell, branch pipe or O-ring has a crease or breakage, replace it.
If not replaced, it will cause a leak.
No.
Name
Comp. No.
Qty
Remark
1
Phillips screwdriver (02)
-
1
Tool
2
Phillips screwdriver (01)
1
Accessory
3
Coin
1
Tool
4
Beaker (500 mL)
1
For cleaning
5
10% nitric acid
500 mL
For cleaning
6
Distilled water
1 L
For cleaning
7
Alcohol (ethanol etc.)
100 mL
For cleaning
8
Cotton waste
1 box
For cleaning
9
Rubber glove
1 set
For cleaning