Springs adjustment of the double pantograph arm
Do not intervene personally in the arm’s settings, have this operation made exclusively by
specialized personnel only.
In the case in which the double pantograph arm does not remain stationary in all
positions wanted by the user, it means that requires some adjustments. The
calibration must be performed as follows: with reference to fig. 11, if it was found that
the section A is spread when the double pantograph tends to go back, it means
that the spring is too tight and in this case we must unscrew slightly the push-spring
nut through the socket wrench supplied. To do this, tilt the section A of about 15 °
and insert the socket wrench into the hole 1 and unscrew slightly.
If on the contrary it is found that stretching the double pantograph, the stretch A
tends to fall forward, it means that the spring must be slightly loaded, then as bifore,
tilt the stretch A of 15 ° and insert the socket wrench into the hole 1 and screw slightly.
This procedure is identical for the section B in which the adjusting key must be
inserted in the hole 2.
fig.11
hole 1
hole 2