MANUAL
Doc. HBE090-3
Installation and User Manual EVOSTYLE N.G.
Mod. MQI006-0
NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095
Grugliasco (TO) Italy Phone and fax + 39.011. 781994 [email protected]
www.newliferadiology.it
11
Efficient grounding is a first necessary condition for the proper functioning of the system,
respecting the symbols indicated for connection:
F – Line (brown) N – Neutral (blue) T – Ground (yellow-green)
Point A grounded power cord; Point B output external lamp
Check carefully all connection cables, plugs and line contacts.
Remember that the suppliers must support a current of at least 10A.
The connection to the external lamps must be done by connecting the output LAMP
indicated with the letter B in the figure above. The lamp must be of value 30W 230-240 Vac.
according to the actual reference standard for plants and locals for medical use (IEC 64-8-
710).