examine. After setting the time, and having moved away to a safe distance from the
radiation source (minimum 2 meters), simply press the button (A) holding it down until the
time on the display starts blinking, the exposure will be accompanied by a beeping sound
and a visual indicator on the timer panel (the led (B) yellow will be lit during the exposure
time).
Once the exposure is finished, must wait until the timer runs out the pause time to make the
next radiography (the timer is paused until the display will flash the time of the last exposure
made).
If the timer is left on without performing any activity for about 15 minutes, it will be on
standby. In this mode, the display will be turned off and only the adult patient led will be
turned on. To disable the standby mode, simply select any button.
After making an exposure, it is possible to know an estimate of the dose emitted expressed
in [mGy] to 20 cm from the focus, by pressing this value will remain visible on the
display for about 3 seconds, after which the display will be displayed again the last
successful exposure time, in this conditions the timer is ready for a new exposure.