MANUAL
Doc. HBE090-3
Installation and User Manual EVOSTYLE N.G.
Mod. MQI006-0
NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095
Grugliasco (TO) Italy Phone and fax + 39.011. 781994 [email protected]
www.newliferadiology.it
10
Information for installation
PLASTIC MATERIAL PLUGS NOT ALLOWED
Utensils and tools required for installation (not supplied)
1 multi meter
1 meter
1 fixed key from 13 mm
1 nandle spanner from 5,5 mm
1 set of Allen spanner
1 spirit level
1 plastic mallet
1 percussion drill with points from Ø 3 to Ø 13 mm
1 thin screwdriver for electrical connection
1 medium screwdriver
1 net feeding cable with three wires (2 cond 1 ground) of section 1,5 mmq for
a reduced length of not over 40 m (for superior lengths use the section cable
corresponding to that indicated from the Country actual standards).
Electrical indications
All works for the feeding electric plant must be performed in conformance with the
actual reference standards for plants and locals for medical use. (CEI 64-8-710).
A feeding of 230-240 V to 50 Hz. is required. For the phase conductors, neutral and
ground, the minimum section must be of 1,5 mmq of copper. It recalls the necessity to
perform the ground connection as required by current legislation.
The plant must be performed and tested by qualified personnel.
The warranty excludes damages caused by erroneous connection.
Wiring and connection conductors sections
A three conductors cable (3x1,5 mmq) is expected between the net switch and the
equipment. For the assembled versions and in all cases of separate installation from the
control unit, a second cable, always 3 x 1.5 sq mm, shall be provided between the
control unit and the support. Please note that the radiological groups are supplied
without a plug for leaving the buyer free to connect to equipment already on site.