Preventive maintenance
CARRIED OUT BY THE DENTAL LABORATORY OPERATOR, WITH ANNUAL CADENCE
Informations for programming routine maintenance on the plant are provided to ensure the
best conditions of use by fully exploiting the diagnostic potential, without compromising any
safety and reliability aspect that characterize the Company.
The verifications and controls are divided into blocks and the operator before starting the
procedures of controls must ensure the availability of the instruction manual:
Radiogenic head monoblock:
plate integrity with the identification data;
Ionizing radiations warning signals integrity;
oil leaks absence verification;
complex integrity: covers shells, spacer cone attachment, rotation system;
monoblock’s correct anchoring verification at the pantograph;
verify the correct 360 ° rotation of the monoblock;
Wall and column support,and pantograph:
verify the correct fixing of the wall support (only for wall version);
move the pantograph in all directions to verify the stability and bilance;
verify that the movement of the complex is fluid, flexible, without obstacles and that it is not
hardening;
check for drawings from fluidifying liquid;