Installation and User Manual EVOSTYLE N.G.
Mod. MQI006-0
NEW LIFE RADIOLOGY S.R.L. Corso Giuseppe Canonico Allamano 13/15 INT.G - 10095
Grugliasco (TO) Italy Phone and fax + 39.011. 781994 [email protected]
www.newliferadiology.it
3
Timer
The timer unit EVOSTYLE N.G. is integrate in
the wall plate; it allows the management
of exposure times and at the same time
guarantees maximum safety in the use of
x-ray tubes for electro purposes for
intraoral diagnostics;
The timer operation is managed by means
of the button with wire. The x-ray delivery is
managed through consent to “dead
man’s switch” so as to ensure maximum
security both for the operator and for the
patient;
The timer of the control panel is provided
with membrane digital keys in order to
facilitate the use;
The timer does not allow the use of
radiographic in fluoroscopy;
The time scale set with the factory default
values complies with R10 scale relative to
the norm EN60601-2-7 (IEC 60601-2-7);
The exposure times displayed on the timer
display are express in ms.
The maximum time that can be set is 3 s
(the display will read 3000) while the
minimum time is 40 ms. (the display will
read 0040);