Monoblock installation
To install the monoblock refer to the procedure described previously for the wall
version from page 20.
To adjust the springs of the double pantograph arm refer the procedures for the wall
version on page 24.
Timer operation
The timer of the device allows the management of the exposure times of the
monoblock at the same time guarantees maximum safety in the use of X-ray tubes
for electro medical purposes for intraoral diagnostic;
The timer control panel is provided with membrane digital buttons so as to facilitate
its use;
The timer functioning is managed by button with spiral cord
to dead man’s device in
such a way as to ensure maximum safety for both the operator and patient;
The timer does not allow the use of X-ray in scopy;
The time range set by the factory values complies with R10 range relative to
standard EN60601-2-7 (IEC 60601-2-7);
The exposure time displayed on the timer are expressed in ms;
The maximum time that can be set is 3 s (the display will read 3000) while the
minimum time is 40 ms. (the display will read 0040);