Wall plate installation
This device is provided with a plaque for the wall mounting with built-in timer
(
A
)
,
closed with a plastic cover (fig.1).
fig. 1
fig. 2
Define the position on the wall of the WALL PLATE
(
A
)
with respect to the amplitude
of the working field chosen and to the size of the structure that the same can assume
right or left of the axis of the wall plate when it is inactive.
Plot on the wall the position of the six holes making sure to check the perpendicularity
with a plumb line. If the electric plant is recessed, track also the corresponding hole
(Fig. 2)
Drill six holes on the wall starting with the tip of
7, enlarging gradually. This to not
demolish the stability and maintain under control the interaxes. For walls of full or
holed or cement bricks use metallic plugs preferable of
12 equipped with
unmissable female grain and separated screw
6 with hexagonal head and washer.
For types of wall of insufficient reliability it is necessary to resort to the construction of a
reinforcement to be defined case by case.
Apply the wall plate and bring it closer parallel to the wall by tightening the 6 screws
alternately, if the wall is not perfectly flat put adequate thicknesses so as not to