46
Measurement of the filament current
The measurement of the filament current can be carried out through the use of an
oscilloscope or by means of a digital multimeter. To perform this measurement, proceed as
follows:
After removing the back cover, referring to Fig.25 Unsolder the wire from the
connector at point A and connect in series a resistance from 1KΩ
fig. 25