40
ERROR TABLE
The errors are displayed with "ERROR:" followed by an indication of the error:
ERROR CODE
CAUSE
SOLUTION
Err1
The
(
A
)
button turns out to
be pressed while pressing
the system power button
Turn off and turn on again
the device;
Should such problem
persist, it means that the
button is blocked in the
“normally closed” mode.
Substitute the button;
Err2
The
(
A
)
button is released
prior to the end of the
exposure time
In order to get out of the
error press a button from
the keyboard at the end of
the cool in procedure;
Err3
The
(
A
)
button is pressed
by the user at the end of
an exposure and
subsequent cooling action
In order to get out of the
error release the button
key;